Moving Sample Volume
To move the Sample Volume position in the D Only state, the system calculates PRF values and the
Phantom Gate position, and updates them on the PW Spectral Doppler image. HPRF is terminated
when HPRF cannot be activated.
When Sample Volume is moved in the 2D Only state, the PRF values don’t change.
NOTE: HPRF is not activated in Simultaneous Mode. It is also not activated if [PRF*2] exceeds 23KHz.
CAUTION:
X
The Phantom Gate position can be located outside the 2D image area in Zoom Mode.
X
Make sure that sample volume and Phantom Gate are not placed together in the measuring
area. If more than two Sample Volumes are located in the vessels, all Doppler components will
appear in the spectrum, causing noise.
PW Spectral Doppler Mode Soft Menu
SV Size / Angle
X
SV Size
Set the size of the Sample Volume. Select a size between 0.5 and 15 by rotating the
Soft Menu
dial-
button
[2]
.
X
Angle
If you press the
Soft Menu
dial-button
[2]
, the angle changes – 60, 0 or 60 degrees. The speed can be
measured accurately by adjusting the angle of the Sample Volume.
Auto Calc
Press the
Soft Menu
dial-button
[3]
to turn it on or o.
If it is turned on, Doppler Trace is performed and the resulting values are displayed. When it is on, the
Mean Trace
button is activated. The displayed values are as below. For information on setting display
values, please refer to "Chapter 7. Utility.”