Dushyant Patel LC-MS/MS Method
S.K.P.C.P.E.R., Kherva 38 M .Pharm. Thesis
10. TRIAL-
10
Symmetry C8,
150 × 4.6 mm,
5 µ
Methanol :
5 mM Amm.
formate
(80:20)
Aqueous
MQC &
External
spiked
MQC
peak shape was
good, ion
suppression was not
observed
11. TRIAL-
11
Symmetry C8,
150 × 4.6 mm,
5 µ
Methanol :
5 mM Amm.
formate
(80:20)
Linearity
& P&A
(Waters
Cartridge)
Within acceptance
criteria
12. TRIAL-
12
Symmetry C8,
150 × 4.6 mm,
5 µ
Methanol :
5 mM Amm.
formate
(80:20)
Linearity
& P&A
(Orochem
Cartridge)
Within acceptance
criteria
4.2.9 METHOD SUMMARY:
Biological matrix : Human plasma
Sample volume required : 0.1 ml
Anticoagulant : Heparin
Analyte : Gabapentin
Internal standard : Levetiracetam
Calibration curve range : GBP : 0.05 µg/ml to 6.00 µg/ml
Detection mode : Mass spectrometry
Analytical technique : Liquid chromatography- Mass
spectrometry
Sample treatment : Solid phase extraction
Quantitation method : Peak area ratio
Weighting factor for Calibration curve : 1/X
2
4.2.10 CHROMATOGRAPHIC CONDITIONS:
Chromatographic
mode
: Reversed phase
Isocratic/gradient
mode
: Isocratic
Tray temperature : 5°C
Injection volume : 5.0 µl
Column : Brand : symmetry
: Type : C8
Vista de página 49