9-38
Operation Manual
Cleaning and Disinfecting Biopsy Kit
Wash and disinfect the biopsy kit to reduce pathogens to the level of 10
-6
. Some components of the
biopsy kit may be disposable. Please read the biopsy kit user manual carefully before use.
Refer to the table of ‘Cleaning and Disinfecting the Probe section in this chapter for detail.
WARNING: Always use protective eyewear and gloves when cleaning and disinfecting biopsy kit.
Cleaning and disinfection of stainless steel biopsy kit
Cleaning
1. After use, remove the biopsy kit from the probe.
2. Disassemble the biopsy kit into its component parts, if applicable.
3. Using a small brush and water, scrub each part to remove trapped material from the biopsy kit.
4. Rinse with water to remove remaining particulates.
Disinfection
1. Disinfect the adapter by autoclaving (Steam) or using gas (Ethylene Oxide).
2. After disinfection, follow the proper post-disinfection procedure for the disinfection method used.
(Please refer to the disinfection user manual, etc.)
3. Inspect the biopsy kit for damage such as cracks, rust or breakage. If damage is evident, discontinue
use of the biopsy kit and contact Samsung Medison Customer Service Department.
Cleaning and Disinfection of plastic biopsy kit
Cleaning
1. After use, remove the biopsy kit from the probe.
2. Disassemble the biopsy kit into its component parts, if applicable. Discard the single-use parts. These
parts cannot be disinfected.
3. Using a small brush and water, scrub each part to remove trapped material from the reusable
components.
4. Rinse with water to remove remaining particulates.