8-6
Operation Manual
Fuse Replacement
The power protection fuse protects the product from excess current. If the power monitoring protection circuit
detects excess current, it shuts o the current to the equipment in order to prevent overheating and to restrict
the ultrasound power output.
If the fuse blows, replace it as shown below.
DANGER:
To avoid risk of electric shock, always disconnect the plug from the system prior to fuse
replacement.
[Figure 8.1 Fuse replacement]
1.
Set this to o by pressing the power switch (See
).
2.
Disconnect the power cord from the wall outlet.
3.
Press down on the middle part (See
) of the used fuse and pull out.
4.
Fit the new fuse into the fuse holder, and press the fuse in to complete replacing the fuse.
Fuse information is shown in the following table.
Input Ratings Fuse Ratings Company Order No.
100-120VAC 10AH/250V Orisel 55T210000
200-240VAC 10A/250V Orisel 55T210000