7-70
Operation Manual
Position / Loop
X Position: Use the
Soft Menu
dial-button
[5]
Position
to set the position of ECG. When its number is
large, ECG will be displayed in upper side of screen. In the Loop mode, press the dial-button to move
to Loop and its setup.
X Loop: Use the
Soft Menu
dial-button
[5]
Loop
to set the ECG area to display. Press the dial-button to
select Loop or 2D.
Gain / Auto Gain
X Gain: Use the
Soft Menu
dial-button
[6]
Gain
to set the ECG gain. Turn the dial-button to set a value
from 10 to 100.
X Auto Gain: When the
Soft Menu
dial-button
[6]
Auto Gain
is pressed, the ECG gain is set automatically.
CAUTION:
X
If the ECG is less then 30Hz, the Heart Rate (HR) is not displayed.
X
In CW Mode, when ECG is active, the error ratio of the Heart Rate (HR) should be within 2%.
[Figure 7.39 ECG]