9-22
Operation Manual
Electric Shocks
The probe uses electrical energy. If it touches conductive materials, there are risks of electric shocks to the
patient or the user.
WARNING:
X
Regularly receive short-circuit examination from the Samsung Medison Customer Support
Department.
X
Do not immerse the probe into liquid.
X
Do not drop the probe or apply mechanical shocks.
X
Inspect the housing, strain relief, lens and seal for damage, and check for any functional problem
before and after each use.
X
Do not apply excessive force to twist, pull or bend the probe cable. It may result in a short circuit.
X
The power protection fuse protects the probe and the product from excess current. If the power
monitoring protection circuit detects excess current, it immediately shuts off the current to the
probe in order to prevent the probe surface from overheating and to restrict the ultrasound
power output.
X
The temperature of the product for making contact with patients is limited under 43°C. The
ultrasound power output (AP&I) is in compliance with US FDA standards.