Exiting PW Spectral Doppler Mode
Press the PW dial-button while the Doppler image is being displayed. You can also press the 2D
dial-button to exit PW Spectral Doppler Mode.
Press the Set button on the control panel to obtain a spectral doppler image.
NOTE: The doppler image can only be obtained in the D Only or Simultaneous states.
PW Spectral Doppler Mode Screen Sample Volume
The doppler spectrum is displayed when the Sample Volume is located above the blood ow
on the 2D image. The size of the Sample Volume is displayed in mm units. Use the trackball to
adjust the position of the Sample Volume.
X
Moving Sample Volume: Use the trackball on the control panel.
X
Resizing Sample Volume: Using the touch screen: Select the SV Size dial-button on the
touch screen.
X
Adjusting Sample Volume Angle: Rotate the Angle dial-button on the control panel to select
a value between -80-80 degrees. Press the Angle dial-button to select either -60, 0, or 60
degrees.
Adjusting Doppler Baseline
Adjust the Baseline by rotating the dial-button on the touch screen.
HPRF (High PRF) Function
Adjust the blood ow above the speed limit at a specic depth. The scale is increased; this can
only be used in PW Spectral Doppler mode (D Only).
NOTE: Setup > Imaging > Common > HPRF
X
Activating HPRF
Increasing the scale at a specic depth to a certain point automatically activates HPRF.
The Phantom Gate will appear on the D Line at a position higher than the sample volume.
Once HPRF starts, PRF does not increase even if you increase the scale value.