Chapter 8 Measurements and Calculations
Optimization of Measurement Accuracy
2D Mode
X
Resolution is in proportion to the frequency of the probe.
X
Penetration is inversely proportional to the frequency of the probe.
X
The highest resolution can be obtained at the focus of the probe where the ultrasound beam is
narrowest.
X
The most accurate measurements can be obtained at the focus depth. As the distance from the
focus increases, the beam width increases, which results in lower accuracy.
X
Using the zoom function or minimizing the depth display makes distance or area
measurements more accurate.
M Mode
X
The accuracy of time measurements can be increased by setting the sweep velocity and the
display format to higher values.
X
The accuracy of distance measurements can be increased by setting the display format to a
higher value.
Doppler Mode
X
Using lower frequency ultrasound is recommended for measurement of faster blood ows.
X
The size of the sample volume is limited by the axial direction of the ultrasound.
X
Using lower frequency ultrasound increases penetration.
X
The accuracy of time measurements can be increased when the sweep velocity is increased.
X
The accuracy of velocity measurements can be increased when the vertical scale is set to
smaller values.
X
It is most important to use an optimal Doppler angle to enhance the accuracy of velocity
measurements.