7-18
User Manual
Timer 1/Timer 2
Starts or resets the timer.
Press the button to start the timer. When the button is re-tapped, the timer stops and the time
is reset. If you don’t tap the button again, the timer will keep running even when the CEUS+
On/O state is changed.
Dual Live
Tap Dual Live on the touch screen and select On or O. When ‘On’ is selected, Dual Live mode
is activated to display both Contrast and 2D images side-by-side. When ‘O’ is selected, Single
mode is activated.
Ref. Line
This option displays the Ref. Line on an image. Select On or O by tapping Ref. Line on the
touch screen. The Ref. Line in CEUS+ mode is used to indicate the same location on the Agent
and Tissue images in Dual Live mode.
Contrast
Select a CEUS+ type. Tap the button on the touch screen to select either Contrast or 2D.
X
Contrast: It displays contrast-applied images.
X
2D: Displays image in 2D Mode.
Flash
Tap the button on the touch screen and select On or O. For the dened frames in Frame,
brighter images are produced.
Flash Frame
Set a value between 3 and 100 frames to specify the period of time during which the Flash
function will operate.
MI Control
Select a value for the mechanical index by tapping the touch screen button or by using the Soft
Menu dial-button on the control panel.
h