8-20
User Manual
VolumeFlow(D)
This is a basic measurement that is only available in Spectral Doppler mode and 2D Image Mode.
Specify two bars on a Spectral Doppler image, and specify two points in 2D Image mode, to
calculate the VolumeFlow.
1. Select VolumeFlow(D) on the monitor or touch screen.
2. Use the trackball and the Set button on the control panel to select two bars. This measurement
is taken in the same way as ‘M Time. This can only be done in the Loop area.
3. Use the trackball and the Set button on the control panel to select two points. 2D Image Mode
will activate automatically so that you can measure the distance. This measurement is taken in
the same way as ‘Distance’.
4. When the measurement has been taken, the result is shown on the screen.
Time(D)
This is a basic measurement that is only available in Spectral Doppler mode. Place two bars on a
Spectral Doppler image to calculate the time between the bars.
1. Tap D Time on the touch screen.
2. Set two bars on the spectrum with the trackball and the Set button.
3. When the measurement has been taken, the result is shown on the screen.
HR(D)
This is a basic measurement that is only available in Spectral Doppler mode. Place two bars on an
image and measure the heart rate between the two bars.
1. Select HR(D) on the monitor or touch screen.
2. Use the trackball and the Set button on the control panel to select two bars.
3. When the measurement has been taken, the result is shown on the screen.