4-4
Operation Manual
Basic Use
The items that can be used commonly in each diagnosis mode are shown below:
QuickScan / Gain
X
QuickScan: The ‘Q Scan mark will appear at the top of an image. In 2D Mode, it is used to
optimize the contrast and brightness of an image by adjusting Gain and TGC automatically. In
PW Spectral Doppler Mode, it is used to optimize the spectrum by adjusting Scale and Baseline
automatically. Press the Exit button to quit the Quick Scan Mode.
X
Gain: Use the QuickScan / Gain dial-button on the control panel. You can adjust the brightness
of an image. If you rotate the Gain dial-button clockwise, its value increases.
TGC (Time Gain Compensation)
Use the TGC slide on the control panel.
In general, ultrasound penetration gets weaker with depth. TGC can be used to compensate for
this eect.
The product provides eight TGC slides for varying depths, allowing you to adjust Gain by area.
Among the eight slides, the top slide represents the shallowest area, while the lower slides represent
the deeper ones.
Move the slide to the right (+) to increase Gain, brightening the image.
CAUTION: Too large a difference in the gain value settings of two adjacent slides may lead to
inaccurate image generation.
Focus
Use the Focus switch on the control panel. Press the upper button to position the focus shallower
or press the lower button to position it deeper.
Depth
Use the Depth switch on the control panel. Press the upper button to make it shallow or press the
lower button to make it deeper. Adjustment range varies depending on the probe type selected.