Vista de página 106
4-23
Chapter 4 Diagnosis Modes
PW Spectral Doppler Mode Screen
On the Spectral Doppler Mode image, the horizontal axis represents time, and the vertical axis
represents speed (or frequency).
Sample Volume
When Sample Volume is positioned over the blood ow in a 2D image, the Doppler Spectrum is
displayed. Information on the size and depth of Sample Volume is displayed in [mm] units. The
position is displayed in the ‘xx.x@yy.y mm’ format. This means that a Sample Volume of size ‘xx.x mm
is located ‘yy.y mm deep. For example, [email protected] means that a Sample Volume of size 2.0 mm
is located 16.7 mm deep.
Use the Change button to adjust the position and the size of the Sample Volume. Each time you
press the Change button, the current status of the Sample Volume will be displayed on the lower
left side of the screen.
X
SV Pos.: You can change the position of the Sample Volume. Use the trackball to move the
Sample Volume position.
X
SV Size: You can change the size of the Sample Volume. Use the trackball to resize the Sample
Volume and press the Set button.
HPRF
This function measures blood ow for which the speed exceeds the specied limit at a specied
depth. It expands the scale into double the size of the original scale. This function is available only
in PW Spectral Doppler Mode (PW Only). HPRF stands for High Pulse Repetition Frequency.
NOTE: To use HPRF function, select ‘On in the Seup > System > Display tab.
X
Running HPRF
If you keep increasing the scale value at a certain depth, HPRF will be activated. The Phantom
Gate will appear on the D Line at a position above the Sample Volume. Once HPRF starts, PRF
does not increase even if you increase the scale value.
X
Exiting HPRF
While HPRF is in use, decrease the scale value by one step to exit HPRF. The PRF value at this
point will be the maximum value in PW Spectral Doppler Mode.