2-14
Operation Manual
Power Connection Part
The power connection part is located at the back of the system.
1
6
2
3
5
4
[Figure 2.7 Power Connection Part]
Item Descriptions
1
Power switch: Supplies or blocks power to the entire system.
2
Power inlet: For the power cable to connect to external power
3
Fuse holder: For the inlet fuse
4
Equipotential terminal: Should be connected to the equipotential bonding network in medical rooms.
5
INPUT/OUTPUT Voltage Selector: Selects the input/output power.
6
Auxiliary socket outlet: Supplies an external peripheral device with power from the product's internal
power supply
NOTE: Cover the power inlet area with a plug bracket and secure it with screws.