Samsung SONOACE R3 Manual do Utilizador Página 267

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 291
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 266
Chapter 9 Probes
9-17
Product Manufacturer or Distributor Telephone number
Metricide CIVCO Co.
+1-800-445-6741(USA)
+1-319-656-4447(Worldwide)
Metrizyme Metrex Research Corp. +1-800-841-1428(USA)
Milton Product & Gamble Australia Pty. Ltd. +61-1800-028-280(Australia)
Nuclean Nation Diagonostics Co.
+1-800-526-3867(USA)
+44(0)-148-264-6020(UK)
Omnicide Cottrell Ltd. +1-800-843-3343 (USA)
Sani-cloth PDI Nice/Pak Products Co. +1-914-365-1602(USA)
Sekusept Extra Henkel Hygiene GmbH. +49-0211-797-0(Germany)
Sporox II Sultan Chemist Inc. +1-800-637-8582(USA)
T-Spray CIVCO Co.
+1-800-445-6741(USA)
+1-319-656-4447(Worldwide)
Virkon Antec International LTD. +1-403-286-1771(USA)
Wavicide Wave Energy System Inc. +1-800-252-1125(USA)
Cleaning
Cleaning is an important procedure that is carried out before disinfecting the probe. The probe must be
cleaned after each use.
CAUTION:
X
Do not use a surgical brush when cleaning probes. The use of even soft brushes can damage the
probe.
X
During cleaning and disinfection, keep the parts of the probe that must remain dry higher than
the other parts during wetting until all parts are dry. This will help prevent liquid from entering
non-liquid-tight areas of the probe.
1. Disconnect the probe from the system.
2. Remove any biopsy adapters or biopsy needle guides. (Biopsy adapters are re-usable and can be
disinfected).
3. Discard sheaths. (Sheaths are single-use items).
4. Use a soft cloth lightly dampened with mild soap or compatible cleaning solution to remove any
particulate matter and body uid that remain on the probe or cable.
Vista de página 266
1 2 ... 262 263 264 265 266 267 268 269 270 271 272 ... 290 291

Comentários a estes Manuais

Sem comentários