5-4
Doppler Signal Adjustment
▶
During velocity measurement, an error may occur depending on the cosine angle between the
blood ow and the ultrasound beam.
▶
For Doppler velocity measurements, the most accurate results can be ensured when the ultrasound
beam is aligned in parallel with the blood ow.
▶
If that is not possible, the angle between them should be adjusted by using the Angle option.
Aliasing
▶
PW Spectral Doppler Mode uses a signal sampling technique to calculate the frequency (or velocity)
spectrum.
▶
Adjust the baseline or the velocity scale to minimize aliasing. A lower frequency probe can also be
used to reduce aliasing.
Calculation Equation
▶
Some of the calculation equations used for clinical purposes originate from hypotheses and approxi-
mation.
▶
All calculation equations are based on medical reports and articles.
Human Error
▶
Human error may occur due to inappropriate use or lack of experience.
▶
This can be minimized through compliance with and thorough understanding of the manuals.