5 - 2 SA8000 SE Operation manual
Measurement Range and Accuracy
NOTE
For optimal performance, the accuracy check should be carried out every year
according to the measurement accuracy table that is provided in the manual.
The measurements provided by the system do not define specific physiological or anatomic
parameters. Rather, what is provided is a measurement of a physical property such as a
distance or velocity for evaluation by the clinician.
For each of the measurements available on the system, the measurement accuracy and range
over which that accuracy is valid are shown in the following tables. Measurement accuracy is
also constrained by the caliper placement capability limit.
The RMI413 phantom is used for all 2D and M Mode measurement accuracy tests. The
RMI1425A phantom is used for Doppler measurement accuracy tests.
Unless specified otherwise by a probe or application, all linear distance movements have the
following accuracy limits.
2D Mode
Measurement Type Range Accuracy Notes
Axial Distance 0.01-25.00 cm +/- 2% or 2 mm
Lateral Distance 0.01-35.00 cm +/- 2% or 2 mm
Diagonal Distance 0.01-25.00 cm +/- 2% or 2 mm
Area 0.01-1,000cm
2
+/- 4% or 0.25 cm
2
Circumstance 0.03-10,000 cm +/- 3% or 5 mm
Volume 0.01-39,000.00 cm
3
+/- 8%
NOTE
1. The accuracies are calculated using the following equations:
Distance error (2% or 2mm) = Image Pixel error (1% or 1mm) + Hardware error
(1% or 1mm)
Area error (4%)= Distance 1 x Distance 2
Volume error (8%)= Distance 1 x Distance 2 x Distance 3
Circumference error (3% or 5mm)= Distance error ( 2% or 2mm) + Calculation
precision ( 1% or 3mm)
2. The phantom used is Multi-purpose phantom, RMI 413, from GAMMAX, Inc.
USA.