4-22
Operation Manual
[Figure 4.11 PW Spectral Doppler Mode - More]
Entering & Exiting PW Spectral Doppler Mode
Tap the PW button on the touch screen. Tap the Update button on the touch screen to obtain a
spectral Doppler image.
NOTE: The Doppler image can only be obtained in the D Only or Simultaneous states.
Sample Volume for PW Spectral Doppler Mode Screen
The Doppler spectrum is displayed when the Sample Volume is located above the blood ow on
the 2D image. The size of the Sample Volume is displayed in mm units. Select the sample volume
with your nger to adjust its position.
X
Moving Sample Volume: Select the sample volume of the study image to adjust its position.
X
Adjusting Sample Volume Angle: Tap the Angle buttons ‘+/-‘ buttons on the touch screen to
select an angle between -80 and 80 degrees. Alternatively, tap the Quick Angle button on the
touch screen to select a preset angle of -60, 0, or 60 degrees.
X
Resizing Sample Volume: Tap the '+/-' button of SV Size on the touch screen to select a setting
between 0.5mm and 15mm. Note that dierent probes provide dierent selectable ranges.