4-22
Operation Manual
[Figure 4.11 PW Spectral Doppler Mode - More]
Entering & Exiting PW Spectral Doppler Mode
Tap the PW button on the touch screen. Tap the Update button on the touch screen to obtain a
spectral Doppler image.
NOTE: The Doppler image can only be obtained in the D Only or Simultaneous states.
Sample Volume for PW Spectral Doppler Mode Screen
The Doppler spectrum is displayed when the Sample Volume is located above the blood ow on
the 2D image. The size of the Sample Volume is displayed in mm units. Select the sample volume
with your nger to adjust its position.
X
Moving Sample Volume: Select the sample volume of the study image to adjust its position.
X
Adjusting Sample Volume Angle: Tap the Angle button’s ‘+/-‘ buttons on the touch screen to
select an angle between -80 and 80 degrees. Alternatively, tap the Quick Angle button on the
touch screen to select a preset angle of -60, 0, or 60 degrees.
X
Resizing Sample Volume: Tap the '+/-' button of SV Size on the touch screen to select a setting
between 0.5mm and 15mm. Note that dierent probes provide dierent selectable ranges.